|
Wescor Thermocouple Psychrometer |
|||
|
The psychrometer can be used to measure total plant water potential (MPa) on fresh leaf tissue, or plant osmotic potential (MPa) using frozen/thawed tissue or expressed sap. Samples (tissue or solutions) equilibrate with water vapor in the chamber atmosphere. Evaporation of a water film on the psychrometer thermocouple cools a thermocouple and generates a negative micro-voltage. The recorded micro-voltage is proportional to the water potential of the sample |
|||
| 1) Set-out the Wescor chambers in a temperature-stable area in the laboratory so their temperature can equilibrate. Minimise handling to avoid temperature effects. | ![]() |
||
| 2) Set-up the meter reader (HR-33T Dew Point Microvoltmeter). |
![]()
|
||
| 3) Collect your plant material from the greenhouse. The plant shoot is harvested and put in a plastic bag or wrapped on saranwrap to prevent desiccation. Speed is essential to reduce measurement bias. | |||
|
4) For total leaf water potential, fresh leaf tissue is used For osmotic potential,
a fully-expanded leaf is taken from each plant, folded and put in
a garlic press. In the press cells are crushed and broken; bean tends
to be extruded as a puree, but corn will produce a clean sap sample.
(an alternative is to rupture cell walls by freezing the sample with
liquid N) It is important to remove debris and sap from the
rim and outside of the sample cup before it is put in the sample changer. |
|||
|
5) Use the mid-size sample chambers.
Using the leaf punch, take several discs from a 'standard' part of
the leaf. Place the discs into the sample well, slide the sample into
the chamber, tighten the chamber knob (do not overtighten).
Allow the sample 10-30 mins to equilibrate with the atmosphere of
the chamber. Avoid handling the chamber to prevent temperature dis-equilibrium.
|
![]() |
||
|
6) Plug the chamber wire into the meter reader 7) Calibrate the micro-voltmeter The Wescor has different calibrations depending on
the size of sample chamber used. For the medium-sized chambers: |
|||
|
8) turn the 'FUNCTION'
knob to "READ"
|
|
9) turn the 'ZERO OFFSET' knob until the meter needle returns to zero. Use the 'coarse' and 'fine' knobs as necessary | |
|
10) turn the 'FUNCTION' knob to "COOL" (this allows water to condense onto the thermocouple - 5 seconds is usually sufficiant) |
![]() |
||
|
11) turn the 'FUNCTION' knob to "DP" (dew point) as water evaporates (cooling the thermocouple) the mV meter-needle will decrease fairly rapidly, but once evaporation reaches equilibrium with the humidity of the chamber the needle will stabilize - THIS IS THE mV TO RECORD. Once the thermocouple dries the mV meter-needle will return to zero. |
![]() |
||
| 12) repeat steps 8-11 to ensure the reading
is stable 13) calculate the sample water potential (MPa) as : y = mV / -7.5 |
|||
|
Copyright © The Ohio State University |
|||